Use of the Ne-213 Organic Scintillator for the Reducti on of Compton Distribution from Gamma-Ray Spectra

Authors

  • Mahmoud A.Elawi
  • Saad N. Yaqoub
  • Fadi H. Khidhr

DOI:

https://doi.org/10.25130/tjps.v25i5.293

Keywords:

Compton distribution, NE-213 detector, γ-ray

Abstract

A method of reducing the Compton distribution from gamma-ray spectra recorded by Nal(Tl) detector using the NE-213 organic scintillator is presented. The similarity of the Compton distributions between both detectors may allow subtracting the NE-213 spectrum from the Nal(Tl) spectrum after making certain calibrations in energy, resolution and count rate at the Compton peak. The method is applied by using the ¹³⁷Cs and ⁶⁰Co gamma-ray   standard sources. The "photopeak" spectra obtained showed that the Compton distribution may be highly reduced and the photopeaks are very well isolated